Sterilisation is monitored in three ways: mechanically, by checking the cycle's time, temperature and pressure readouts; chemically, with indicators such as autoclave tape that only confirm exposure, or integrators that respond to several parameters; and biologically, with spores of an organism that is highly resistant to the particular sterilising agent. For steam under pressure the indicator is Geobacillus stearothermophilus, a thermophilic spore-former whose spores are among the most resistant to moist heat. A spore strip or self-contained vial is processed in a normal load, then incubated at 55 to 60 degrees Celsius; no growth means the cycle was lethal. Infection control guidance recommends spore testing at least weekly and with every load that contains implantable devices, and a positive test takes the autoclave out of service until the fault is found. Option B, Bacillus atrophaeus, is the indicator for dry heat and ethylene oxide. Option C, Bacillus pumilus, is used for ionising radiation. Option D is a pathogen, not an indicator, and its spores are readily killed by autoclaving. Exam tip: match the organism to the method, stearothermophilus for steam, atrophaeus for dry heat and gas, pumilus for radiation.